| pH |
| Measuring Range |
-2.00 to 16.00 pH |
n/a |
-2.00 to 20.00 pH |
| Resolution |
0.1/0.01 pH |
n/a |
0.1/0.01 pH |
| Selectable Resolution |
Yes |
n/a |
Yes |
| Accuracy |
± 0.01 pH |
n/a |
± 0.01 pH |
| Pre-Defined Buffer Groups |
3 |
n/a |
5 |
| ORP, Redox |
| Measuring Range |
±2000.0 mV |
n/a |
±2000.0 mV |
| Resolution |
1 mV |
n/a |
0.1mV |
| Accuracy |
± 1 mV |
n/a |
±0.5 mV |
| Units |
mV, RmV |
n/a |
mV, RmV |
| Conductivity |
| Measuring Range |
n/a |
0.001μS/cm to 19.99 μS/cm 20 μS/cm to 199.9 μS/cm 200 μS/cm to 1999 μS/cm 2.00 mS/cm to 19.99 mS/cm 20.0 mS/cm to 1000 mS/cm |
0.01 μS/cm to 19.99 μS/cm 20 μS/cm to 199.9 μS/cm 200 μS/cm to 1999 μS/cm 2.00 mS/cm to 19.99 mS/cm 20.0 mS/cm to 500.0 mS/cm |
| Resolution |
n/a |
0.001 μS/cm minimum; auto-range |
0.01 μS/cm minimum; auto-range |
| Accuracy |
n/a |
± 0.5 % Reading ± 2 Least Signficant Digit |
± 0.5 % Reading ± 2 Least Signficant Digit |
| Reference Temperature |
n/a |
20 °C, 25 °C |
20 °C, 25 °C |
| Cell Constants |
n/a |
0.01 to 10.00 cm-1 |
0.001 to 10.00 cm-1 |
| Temperature Compensation |
n/a |
Linear (0 to 10.0%/ºC), off |
Linear (0 to 10.0%/ºC), off |
| TDS |
| Measuring Range |
n/a |
0.1 mg/L to 200 g/L |
0.1mg/L to 199.9 g/L |
| Resolution |
n/a |
0.01 mg/L minimum, auto-range |
0.01 mg/L minimum, auto-range |
| Accuracy |
n/a |
± 0.5 % Reading ± 2 Least Signficant Digit |
± 0.5 % Reading ± 2 Least Signficant Digit |
| TDS Factor Range |
n/a |
Linear, 0.01 to 10.00, default 0.5 |
Linear, 0.01 to 10.00, default 0.5 |
| Resistivity |
| Measuring Range |
n/a |
1 to 100 MΩ-cm |
2 to 100 MΩ-cm |
| Resolution |
n/a |
0.01 Ω-cm auto-range |
0.01 Ω-cm auto-range |
| Accuracy |
n/a |
± 0.5 % Reading ± 2 Least Signficant Digit |
± 0.5 % Reading ± 2 Least Signficant Digit |
| Practical Salinity |
| Measuring range |
n/a |
0 to 100 psu |
0 to 100 psu |
| Resolution |
n/a |
0.01 psu minimum, auto-range |
0.01 psu minimum, auto-range |
| Accuracy |
n/a |
± 0.5 % Reading ± 2 Least Signficant Digit |
± 0.5 % Reading ± 2 Least Signficant Digit |
| Temperature |
| Measuring Range |
-5.0 to 110.0°C 23.0°F to 230.0°F |
| Resolution |
0.1 °C, 0.1 °F |
| Accuracy |
± 0.5 °C, ±0.5 °F |
± 0.3 °C, ±0.3 °F |
± 0.3 °C, ±0.3 °F |
| Calibration |
No Calibration |
No Calibration |
No Calibration |
| Calibration |
| Calibration points |
Up to 3 points |
1 point cell constant calibration; 6 conductivity standard solutions available (10, 84, 146.5, 500, 1413 μS/cm and 12.88mS/cm) |
Up to 5 points for pH; 1 point cell constant calibration; 6 conductivity standard solutions available (10, 84, 146.5, 500, 1413 μS/cm and 12.88mS/cm) |
| Calibration Sign |
Slope/offset & Face |
Cell Constant & Face |
Slope/offset & Cell Constant & Face |
| Calibration mode |
Linear |
Linear |
Linear |