Digital Benchtop pH-ISE-Conductivity Meter
The HI6522 kit includes the necessary modules and probes to measure pH, temperature and conductivity, providing everything needed to start measuring straight out of the box. This advanced multimeter features WiFi and ethernet connectivity.
This kit is compatible with a wide range of Hanna Instruments electrodes for ISE measurement. For advice on the best electrodes for your needs speak to your Labequip representative.
Together with appropriate electrodes, the meter provides highly accurate and reliable measurement for all laboratories. Up to three measurement channels can be viewed simultaneously. Views can be mixed and matched. Data can be shown in basic or in full detail, in a table or plotted as a graph.
The HI-6000 Series allows for up to 5-point calibration for pH, ISE and Conductivity for optimum results.
- pH calibration up to 5 points with 8 stored standard buffers and 5 custom buffers
- ISE calibration up to 5 points with 5 stored standards and 5 custom buffers
- Conductivity calibration up to 5 points with 6 stored standards and 1 user standard
- Data can be shared via Wi-Fi, Ethernet, email, web server, USB flash drive or PC
- No software is required
- Conveniently connect a keyboard, a barcode scanner, a printer or two stirrers to improve your workflow
- User-friendly icons and symbols enable easy navigation and interpretation of instrument functions
- Choice of five different display modes: Standard, Simple GLP with calibration information, Full GLP with electrode status and calibration point details, Interactive graph updated in real-time,
- Tabulated data with date, time and notes
- 10 user profiles - For direct access to daily analyses, users can save their procedures in profiles grouping all their method configuration data.
- Each user profile is password-protected
SPECIFICATIONS:
Range:
pH: -2.0 to 20.0 pH; -2.00 to 20.00 pH; -2.000 to 20.000 pH
Conductivity: 0.000 to 9.999 μS/cm; 10.00 to 99.99 μS/cm; 100.0 to 999.9 μS/cm;
1.000 to 9.999 mS/cm; 10.00 to 99.99 mS/cm; 100.0 to 1000.0 mS/cm
mV: -2000.0 mV to 2000.0 mV
ISE: 1 x 10-6 to 9.99 x 1010 concentration
Temperature: -20.0 to 120.0 °C; -4.0 to 248.0 °F; 253.0 to 393.0 K
Resistivity: 1.0 to 99.9 Ω・cm; 100 to 999 Ω・cm; 1.00 to 9.99 KΩ・cm; 10.0 to 99.9 KΩ・cm; 100 to 999 KΩ・cm; 1.00 to 9.99 MΩ・cm; 10.0 to 100.0 MΩ・cm
Total Dissolved Solids (TDS): 0.000 to 9.999 ppm; 10.00 to 99.99 ppm; 100.0 to 999.9 ppm; 1.000 to 9.999 ppt; 10.00 to 99.99 ppt; 100.0 to 400.0 ppt; actual TDS
Salinity: 0.00 to 42.00 PSU - Practical Scale
0.00 to 80.00 ppt - Natural Sea Water
0.0 to 400.0 % - Percent Scale
Resolution:
pH: 0.1 pH; 0.01 pH; 0.001 pH
Conductivity: 0.001 μS/cm; 0.01 μS/cm; 0.1 μS/cm; 0.001 mS/cm; 0.01 mS/cm; 0.1 mS/cm
mV: 1 mV; 0.1 mV
ISE: 1; 0.1; 0.01; 0.001 concentration
Temperature: 0.1 °C / 0.1 °F / 0.1 K
Resistivity: 0.1 Ω・cm; 1 Ω・cm; 0.01 KΩ・cm; 0.1 KΩ・cm; 1 KΩ・cm; 0.01 MΩ・cm; 0.1 MΩ・cm
Total Dissolved Solids (TDS): 0.001 ppm; 0.01 ppm; 0.1 ppm; 0.001 ppt; 0.01 ppt; 0.1 ppt
Salinity: 0.01 for Practical Scale / Natural Sea Water
0.1 % for Percent Scale
Accuracy:
pH: ±0.1 pH; ±0.01 pH; ±0.002 pH (±1 last significant digit)
Conductivity: ±1 % of reading ±0.010 μS/cm, whichever is greater
mV: ±0.2 mV ±1 last significant digit
ISE: ±0.5% (monovalent ions); ±1% (divalent ions)
Temperature: ±0.2 °C / ±0.4 °F / ±0.2 K
Resistivity: ±1 % of reading or ±1 Ω・cm, whichever is greater
Total Dissolved Solids (TDS): ±1 % of reading or ±0.01 ppm, whichever is greater
Salinity: ±1 % of reading
Temperature Compensation: Automatic or manual
Calibration: Automatic, up to five-point calibration, seven fixed standard solutions available for each measurement unit, and five user defined standards
Resistivity
Range 1.0 to 99.9 Ω・cm; 100 to 999 Ω・cm; 1.00 to 9.99 KΩ・cm; 10.0 to 99.9 KΩ・cm; 100 to 999 KΩ・cm; 1.00 to 9.99 MΩ・cm; 10.0 to 100.0 MΩ・cm
Resolution 0.1 Ω・cm; 1 Ω・cm; 0.01 KΩ・cm; 0.1 KΩ・cm; 1 KΩ・cm; 0.01 MΩ・cm; 0.1 MΩ・cm
Accuracy ±1 % of reading or ±1 Ω・cm, whichever is greater
Total Dissolved Solids (TDS)
Range 0.000 to 9.999 ppm; 10.00 to 99.99 ppm; 100.0 to 999.9 ppm; 1.000 to 9.999 ppt; 10.00 to 99.99 ppt; 100.0 to 400.0 ppt; actual TDS (with 1.00 factor)
Resolution 0.001 ppm; 0.01 ppm; 0.1 ppm; 0.001 ppt; 0.01 ppt; 0.1 ppt
Accuracy ±1 % of reading or ±0.01 ppm, whichever is greater
Salinity
Range 0.00 to 42.00 PSU - Practical Scale
0.00 to 80.00 ppt - Natural Sea Water
0.0 to 400.0 % - Percent Scale
Resolution 0.01 for Practical Scale / Natural Sea Water
0.1 % for Percent Scale
Accuracy ±1 % of reading
Calibration 1 point for Percent Scale, using 100% salinity calibration solution
Temperature
Range −20.0 to 120.0 °C; -4.0 to 248.0˚F; 253.0 to 393.0 K
Resolution 0.1 °C; 0.1 °F; 0.1 K
Accuracy ±0.2 °C; ±0.4 °F; ±0.2 K
Calibration Single point, adjustable
| Stock Number | N2485 |
| Model | HI 6522 |
| Manufacturer | Hanna Instruments |
| Product Brochure | |