Microplate Reader
|
25862
|
Bio-Tek Instruments ELx800MER
|
As a stand-alone microplate reader, the ELx800 is designed for applications within the clinical, biotechnology research and pharmaceutical areas. The on-board data reduction surpasses many PC software packages with its extensive curve-fitting, cutoff calculation, data transformation and validation capabilities. The ELx800 interfaces with BioTek's Gen5 Data Analysis Software for added functionality. FROM MANUFACTURERS LITERATURE:
Features:
Extensive on-board data analysis:
Curve-fitting options including linear, quadratic and 4-P
Control and assay validation with complete results reporting
Transformation formulas
Cutoff and call criteria
Memory:
Stores 55 assay definitions in memory (up to 75 custom assays)
Saves 8 microplate test results
Stores up to 25 standard curves for reuse
Multiple Plate Formats:
384-well plate capability for High Throughput Screening applications
Routine ELISA's and cellular assays in 24-, 48-, and 96-well analysis
Specialized applications in 60-, 72-, and 96-well Terasaki formats
On-Board Diagnostic Testing:
System Test checks the entire optics path
Universal Test Plate software tests for alignment, accuracy, linearity and repeatability
Language versions available: English, French, German, Italian, Portuguese, Spanish
Interfaces to many robotic workstations (optional)
Expandable capabilities with BioTek's Gen5 Data Collection and Analysis Software
SPECIFICATIONS:
Detection method: Absorbance
Read method: Endpoint; Kinetic and well area scanning via computer control
Microplate types: 6-, 12-, 24-, 48-, 96-well plates 384-plates
Software: Gen5 Data Analysis Software (optional)
Onboard software: 55 user-programmable protocols
Light source: Tungsten Halogen
Wavelength selection: Filters
Wavelength range: 400 to 750nm; 340 to 750nm (UV option)
Bandpass: 10 nm
Dynamic range: 0 to 3.0 OD
Resolution: 0.001 OD
Filter wheel capacity: 5 positions
Filters supplied: 4 filters (5 with UV option)
OD accuracy: <1% at 2.0 OD typical
OD linearity: <1% at 2.0 OD typical; <3% at 3.0 OD typical
OD repeatability: <0.5% at 2.0 OD typical
Reading speed: 30 secs single wavelength
|
Used
|
120 Volts, 60 Hz,
|
1
|
|
Thank you for your information request on Labequip
stock #25862, Microplate Reader, Bio-Tek Instruments ELx800MER.
Please contact us for pricing, shipping charges, or a formal quotation...

PHOTOS: These photos are reduced in size for display on the internet. Please
contact us if you wish to receive larger, high-resolution images by e-mail.
We will also be pleased to provide a close-up of any part of the instrument
you specify.
|
|
|
 |
|
|
 |
|
|
 |
|
|
|
|
|