Cell Harvester
|
23416
|
Skatron Micro 96S - 10056
|
For high throughput testing, the 96-well cell harvesters will harvest all 96 samples simultaneously, or harvest 24 wells at a time, selectable by rows. The 96-well cell harvesters have flexible programming to meet the specific needs of a wide variety of applications. Parameters can be individually set to adapt the program to most standard applications: twelve washes from any of three inlets, separation of radioactive waste, timing of the prewet, and washing and drying sequences. To address more demanding assays, the 96-well cell harvesters allow for a number of harvester and wash combinations.
SPECIFICATIONS (to be verified upon sale)
Harvest Form (input): Microplates 8x12
Filter Format Config: 6x16
Filtering Area: 9 mm
Counter Compatibility: Liquid scintillation counters and Wallac Betaplate counter
Wells Harvested Simultaneously 24/48/72/96 programmable
No. Outlets 2 hot/cold
No. Inlets 3
Harvest Programming Fully programmable, 10 user-selectable programs, 12 harvest sequences, soak/pause
Harvest Wash Flow Continuous/pulsed
Internal Air Pressure Source Yes, user-set air pressure 0.1-0.7 bar
|
Used
|
110/220 Volts, 150W, 1A
|
1
|
|
Thank you for your information request on Labequip
stock #23416, Cell Harvester, Skatron Micro 96S - 10056.
Please contact us for pricing, shipping charges, or a formal quotation...

PHOTOS: These photos are reduced in size for display on the internet. Please
contact us if you wish to receive larger, high-resolution images by e-mail.
We will also be pleased to provide a close-up of any part of the instrument
you specify.
|
|
|
 |
|
|
 |
|
|
 |
|
|
 |
|
|
|
|
|